Comparison of the Smart
Fieldmeter®and Smart Fieldmeter ®Digital to portable battery operated
industry leading field strength
measuring instruments
|
Parameter |
Smart Fieldmeter® Digital: RFP-05/PI-01 |
Smart Fieldmeter®: RFP-04/PI-01 |
Brand A |
Brand B |
Brand C |
Brand D |
|
Electric field
probe type |
Isotropic, RMS, 3-axis. Meter supports up to 4 probes. |
Isotropic, RMS, 3-axis. |
Isotropic, RMS, 3-axis, X,Y, Z. |
Isotropic, RMS, 3-axis. |
Isotropic, RMS, 3-axis, X,Y, Z. |
Isotropic, RMS, 3-axis. |
|
Probe mounting |
Mounted on the meter or connected though the cable |
Mounted on the meter or connected though the cable |
Mounted on the meter |
Mounted on the meter |
Mounted on the meter |
Mounted on the meter |
|
Frequency |
0.2 MHz-3 GHz (other variants available) |
0.2 MHz-3 GHz (other variants available) |
0.1 MHz-5 GHz |
0.1 MHz-3 GHz |
0.1 MHz-3 GHz (other probes available) |
0.1 MHz-2.5 GHz 0.1-1 MHz purely indicative. |
|
Frequency
response |
+/- 1 dB (0.5 MHz-1 GHz), <+/- 2.5 dB (1- 3 GHz) |
+/- 1 dB (0.5 MHz-1 GHz), <+/- 2.5 dB (1- 3 GHz) |
+/- 1 dB(1 MHz –4 GHz) |
+/- 1.5 dB (0.1 MHz-1 GHz) <+/- 2.4 dB (1-3 GHz) |
+/- 0.5 dB (10 -300MHz) <+/- 1.5 dB (3 MHz -3 GHz) |
+/- 1.5 dB (1 MHz-1 GHz) <+/- 2 dB (1-2.5 GHz) |
|
Isotropicity |
+/- 0.5 dB (@100 MHz), <+/- 1.5 dB (0.2 MHz-3 GHz) |
+/- 0.5 dB (@100 MHz), <+/- 1.5 dB (0.2 MHz-3 GHz) |
+/- 0.7 dB |
+/- 0.5 dB (probe), +/- 1 dB (probe with meter), |
<+/- 0.8 dB, 0.5 dB ( 930 and 1800 MHz) |
Not specified |
|
Field strength
range |
0.2-500 V/m (standard) 0.4-800 V/m (variant UL) |
0.2-500 V/m (standard) 0.4-800 V/m (variant UL) |
0.4-800 V/m |
0.6-800 V/m |
0.3-300 V/m |
0.1-200 V/m |
|
Ranges |
Auto range |
4 Ranges: 2, 20, 200, 600 V/m |
Auto range |
Auto range |
Auto range |
4 Ranges: 1, 10, 100, 200 V/m |
|
Resolution |
0.01 V/m |
0.01 V/m |
0.01 V/m |
0.01 V/m |
0.01 V/m |
0.1 V/m |
|
Usable
sensitivity |
0.2 V/m |
0.2 V/m |
0.3 V/m |
0.6 V/m |
0.3 V/m |
1 V/m (Accuracy 0.7 V/m) |
|
Calibration
accuracy |
+/- 0.5 dB |
+/- 0.5 dB |
Not specified |
+/- 1 dB |
+/- 0.8 dB |
Not specified |
|
Linearity
accuracy |
< +/- 1.0 dB in the whole range |
< +/- 1.5 dB 10-150% of any range |
Not specified. |
< +/- 3 dB (0.6-1.25 V/m) < +/- 1 dB (1.25-2.5 V/m) < +/- 0.7 dB (>2.5 V/m) |
Not specified. |
< +/- 0.7 V/m (0.1-1 V/m) < +/- 0.5 V/m (1-10 V/m) < +/- 1 dB (10-100 V/m) < +/- 2 dB (100-200 V/m) |
|
Display units |
V/m, nW/cm2, uW/cm2, mW/cm2 |
V/m |
V/m, mW/cm2, W/m2 |
V/m, mW/cm2, W/m2 |
V/m, mW/cm2, uW/cm2, W/m2 |
V/m, uW/cm2 |
|
Time response |
Multimode: Time Averaging 30 msec-30 min (selectable). Space Averaging (manual). Fast/Slow probe response. |
Multimode: Average: 2 sec, Pulse: 0.1 sec, Peak: 50 ms/14 sec. |
Not specified |
1 sec |
0.15-0.9 sec |
0.4 sec |
|
Zeroing |
Auto zero, prgrammable. |
Auto zero |
Auto zero |
Auto zero |
Auto zero |
Auto zero |
|
Data storage |
Internal datalogging on Memory Card, unlimited
storage, Programmable time interval 100 msec-30 min. |
External data logger or PC link kit |
Internal |
External (high end model - internal) |
Internal |
Internal |
|
PC interface |
RS232, wired or optically decoupled. |
External RS232 |
RS232 |
RS232 |
RS232 |
RS232 |
|
Size |
Meter: 130x80x30 mm, Probe: 230x58 mm. |
Meter: 130x80x30 mm, Probe: 230x58 mm. |
Meter: 163x87x32 mm Probe: Not specified. |
Meter with Probe: 465x96x64 mm |
Meter: 240x108x50 mm Probe: 317x58 mm |
Meter: 216x72x37 mm Probe: 320x50 mm |
|
Battery
operating time |
25-100 Hrs. |
100 Hrs. |
20 Hrs. |
8-15 Hrs. |
20 Hrs. |
30 Hrs. |
|
Weight |
0.45 kg. |
0.33 kg. |
0.64 kg. |
0.45 kg. |
1.2 kg. |
Meter only 0.35 kg. |