The difference between “Test Instruments” and “Indicators”

 

There are many digital field meters on the market.

But only a few can be called “Test Instruments”. Smart Fieldmeter (TM) is one of them.

Other field measuring devices are only “Indicators”. The basics for that distinction are explained below.

 

For most test application the test instrument is intended to measure the value of electric component E of the electromagnetic field at a given point.  The signal frequency and orientation of vector E is not typically known.

 

Therefore, good field meter must have the following features:

 

Table

 

Parameter of the

Field Meter

Required Value and Desired Accuracy

Smart Fieldmeter (TM)

Notes

 

Frequency Response of the Probe

Flat in the test range. 

+/- (0.5 – 3) dB

+/- 1 dB (0.2 MHz - 1 GHz),

+/- 2.5 dB (1 - 3 GHz)

Requires probe detached from the meter metal body (to avoid parasitic coupling) and small broadband non-resonant antenna.

 

Isotropicity of the Probe

Isotropic.

+/- (0.5 2.5 dB)

Better than  +/- 1.5 dB,

Typically,  +/- 0.6 dB @100 MHz

Requires probe with 3 (X, Y, Z) orthogonal antennas.

 

Calibration Accuracy

NIST Traceable.

+/- (0.5 – 1) dB

NIST Traceable.

+/- 0.5 dB

NIST traceable calibration guarantees device correct measurements.

 

Field Strength Linearity of the Meter

Linear in the field test range.

+/- (0.5 - 3 dB)

Better than +/- 1.5 dB.

Typically, +/- 1 dB.

Any range 10-150% of full scale,

0.2 - 600 V/m.

Requires linearization technique depending on the type of probe detector is used.

RMS (Root Mean Square) Measurement Range

RMS in whole test range.

+/- (0.5 - 3) dB error in the whole field strength range.

RMS.

<0.5 dB (for fields less than 60 V/m),

<2.3 dB (at 600 V/m)

 

In multi-signal environments the errors may be higher.

 

Response Time

Typically, 0.5 - 2 sec.

Some applications require much faster response.

Three modes of operation:

Average: 2 sec,

Pulse: 0.1 sec,

Peak: 50 ms/14 sec.

Pulse mode is good for close loop field control during the EMC testing.  Peak mode helps to identify the AM modulated signals.

Zeroing

Automatic or user activated even in the presence of strong field.

Automatic or user activated even in the presence of strong field.

Absolute must in the field.

 

 

From the Table above we see that in order to qualify as test instrument, field meter must have at least:

 

- Special detached probe with three small broadband antennas

 - Linear RMS meter with wide dynamic range 

-  NIST traceable calibration

 

 

Based on that, being “digital” doesn’t make good test instrument,

unless major test instrument features are present, so

most of the low cost “Field Meters” on the market are simply “Field Indicators”

because they have one or few of the following drawbacks:

 

-Don’t have the broadband isotropic probe

- Have antenna built into meter body

- Have directional or long telescopic antenna

- Don’t display the RMS value of the field

- Require manual zeroing

- Don’t have (or can’t have!) the NIST traceable calibration

 

 

Smart Fieldmeter (TM) doesn’t have any of those limitations,

and is a true “Test Instrument”.

For detailed comparison of Smart Fieldmeter to some industry leaders,

see Application Note AN-005.